[koji, gautam]
With Koji's help, I got repeatable and reliable DRMI locking going again tonight - this is with the AS path optics for the spectroscopy measurement in place, although the AUX laser remained shuttered tonight. Results + spectra tomorrow, but here's what I did:
- Initial alignment procedure was as usual - use arms+ASS to align ITMs, and then PRMI carrier+ASS to align PRM and BS.
- Found the appropriate gains and demod phases.
- Measured loop TFs - PRCL is a big mystery. Used these to finalize loop gains.
- Ran some sensing lines.
- Whitened DRMI PDs for a calibrated "low-noise" spectrum (although the coils were not de-whitened).
As I have found before, it is significantly easier to get the locking going post 11pm - the wall Seis BLRMS don't look that much quieter at midnight compared to 10pm, but this might be a scaling issue. I'll do a quantitative assessment next time... Also, Foton takes between 25-45 secs to save an updated filter (timed twice today). |