Message ID: 13926     Entry time: Thu Jun 7 14:35:26 2018
Author: keerthana 
Type: Update 
Category: elog 
Subject: Table- useful for doing the scanning. 

I think this table will help us to fix the scanning range of the Marconi frequency. This will also help in predicting the position of the resonance peak corresponding to the injected frequency.

fdiff = fm ±80 MHz ;                     fdiff = N*FSRy ;              FSRy = 3.893 MHz.

N = Integer number fdiff =injected fm = Marconi frequency
1 3.893 76.107
2 7.786 72.214
3 11.679 68.321
4 15.572 64.428
5 19.465 60.535
6 23.34 56.66
7 27.251 52.749
8 31.144 48.856
9 35.037 44.963
10 38.93 41.07
11 42.79 37.21
12 46.716 33.284
13 50.609 29.391