Message ID: 1390     Entry time: Wed Mar 11 22:57:48 2009     In reply to: 1388     Reply to this: 1402
Author: Yoichi 
Type: Update 
Category: Locking 
Subject: Calibrated XARM error signal spectrum 
I did a rough calibration of the XARM error spectrum.
See the attached calibrated spectrum.

I started from this Rana's elog entry.
http://www.ldas-sw.ligo.caltech.edu/ilog/pub/ilog.cgi?group=40m&task=view&date_to_view=04/07/2005&anchor_to_scroll_to=2005:04:07:20:28:36-rana

I first injected a 20Hz sin signal into C1:SUS-ETMX_LSC_EXC and measured the response to the ETMX SUSPOS.
Using the calibration of the SUSPOS given in the above entry, I calibrated the ETMX coil actuation efficiency.
It was 3.4e-12 m/cnt @20Hz for C1:SUS-ETMX_LSC_EXC.

Then I locked the X-arm and injected a calibration peak at 20Hz.
From the ratio of the peaks in C1:SUS-ETMX_LSC_IN2 and C1:LSC-XARM_IN1, I calibrated the X-arm error signal to be 4.2e-13 m/cnt.
We have to also take into account the cavity pole of the arm, 1525Hz (the design value, may not be actual).
So I used the following calibration in the DTT:

G: 4.2e-13
P: 1525
Z:

Note that the attached spectrum shows the actual motion of the X-arm (or equivalent frequency noise) after suppressed by the feedback servo,
unlike conventional noise spectra showing "virtual" displacement which would have been induced in the absence of servos.
Attachment 1: XarmErrorSpeCalibrated.pdf  46 kB  | Hide | Hide all
XarmErrorSpeCalibrated.pdf