Mon Mar 19 23:44:00 2018, gautam, Update, IOO, IMC loop checkup 
|
Tue Mar 20 16:53:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 21:08:03 2018, gautam, Update, IOO, IMC loop checkup
|
Tue Mar 20 22:44:45 2018, gautam, Update, IOO, IMC loop checkup
|
Wed Mar 21 15:52:45 2018, gautam, Update, IOO, MC error point calibration  
|
Wed Mar 21 17:31:21 2018, gautam, Update, IOO, MC error point calibration
|
Wed Mar 21 21:13:44 2018, gautam, Update, IOO, IMC noise budget
|
Tue Mar 27 01:39:44 2018, gautam, Update, IOO, PSL noise eater was off
|
Tue Mar 27 19:32:03 2018, arijit, Update, IOO, PSL noise eater was off 
|
Tue Mar 27 23:37:35 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:31:39 2018, gautam, Update, IOO, MC REFL PD removed
|
Wed Mar 28 21:47:37 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement 
|
Thu Mar 29 17:57:36 2018, arijit, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 06:14:31 2018, rana, Update, IOO, MCREFL_PD Optical response measurement
|
Fri Mar 30 22:37:36 2018, Kevin, Update, IOO, MCREFL_PD Optical response measurement
|
Thu Apr 5 04:36:56 2018, Kevin, Update, IOO, Coil driver noise  
|
Thu Apr 5 12:13:18 2018, Koji, Update, IOO, Coil driver noise
|
Fri Apr 6 21:39:09 2018, gautam, Update, IOO, MC2 suspension health checkup
|
Mon Apr 9 18:46:03 2018, gautam, Update, IOO, Further debugging 
|
Tue Apr 10 14:28:44 2018, gautam, Update, IOO, Further debugging
|
Wed Apr 11 01:34:31 2018, gautam, Update, IOO, Activities today
|
Thu Apr 12 18:12:49 2018, gautam, Update, ALS, NPRO channels hijacked  
|
Fri Apr 13 11:02:41 2018, gautam, Update, ALS, EX fiber polarization drift
|
Fri Apr 13 17:56:26 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 14 14:42:09 2018, gautam, Update, ALS, Fibers switched out
|
Tue Apr 17 14:08:29 2018, gautam, Update, ALS, Fibers switched out
|
Fri Apr 20 00:26:34 2018, gautam, Update, ALS, Fibers switched out
|
Sat Apr 21 20:25:12 2018, gautam, Update, ALS, PSL fiber pickoff status
|
Thu Apr 19 00:03:51 2018, gautam, Update, IOO, More IMC NBing 
|
Thu Apr 19 17:15:35 2018, gautam, Update, IOO, More IMC NBing 
|
Sun May 20 23:43:50 2018, gautam, Update, IOO, Coil driver noise re-measurement    
|
Tue May 22 17:26:25 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Tue May 22 23:28:01 2018, gautam, Update, IOO, MC1 Coil Driver pulled out
|
Wed May 23 17:58:48 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Wed May 30 10:17:46 2018, gautam, Update, IOO, MC1 Coil Driver pulled out 
|
Fri Apr 6 22:23:53 2018, Kevin, Update, IOO, Coil driver noise
|
|
Message ID: 13883
Entry time: Wed May 23 17:58:48 2018
In reply to: 13880
Reply to this: 13896
|
Author: |
gautam |
Type: |
Update |
Category: |
IOO |
Subject: |
MC1 Coil Driver pulled out |
|
|
- Marked up schematic + photo post changes uploaded to DCC page.
- There was a capacitor in the DAQ monitor path making a 8kHz corner that I now removed (since the main point of this front panel HPF monitor point is to facilitate easy coil driver noise debugging, and I wanted to be able to use the SR785 out to high frequencies without accounting for an additional low pass). Transfer function from front panel LEMO input to front panel LEMO monitor is shown in Attachment #1.
- Voltage noise measured at DB25 output (with the help of a breakout cable and SR560 G=100) with front panel LEMO input terminated to 50ohm, Bias input grounded, and pin1 of U21A grounded (i.e. watchdog enabled state) is shown in Attachment #2. This measurement was taken on the electronics bench.
- Inside the lab (i.e. coil driver board plugged into eurocrate), the noise measured in the same way looks identical to what was measured in elog13870.
- I tried repeating the measurement by powering the board using an bench power supply and grounding the bias input voltage near 1X6, and the strange noise profile persists. So this supports the hypothesis that some kind of environmental pickup is causing this noise profile. Needs more investigation.
In any case, if it is indeed true that the optic sees this current noise, the place to make the measurement is probably the Sat. Box. Who knows what the pickup is over the ~15m of cable from 1X6 to the optic.
Quote: |
Detailed report + photos to follow
|
|
|
|
|
|