Entry  Sun Apr 29 22:53:06 2018, gautam, Update, General, DARM actuation estimate darmEst.pdf
    Reply  Tue May 1 19:37:50 2018, gautam, Update, General, DARM actuation estimate darmEst.pdfdarmEst_time.pdf
       Reply  Mon May 7 16:23:06 2018, gautam, Update, General, DARM actuation estimate darmProj.pdf37.pngMCalign_20180507.png
          Reply  Mon May 7 20:01:14 2018, Rorpheus, Update, General, Use anti-dewhitening + show CARMA/DARMA 
             Reply  Thu May 10 14:13:22 2018, gautam, Update, General, More refinement of DARM control signal projection darmProj.pdfdarmOLTF.pdfDWcomparison.pdf
                Reply  Fri May 11 13:58:42 2018, rana, Update, General, More refinement of DARM control signal projection 
                   Reply  Fri May 11 19:02:52 2018, gautam, Update, General, More refinement of DARM control signal projection darmProj.pdfdarmOLTF.pdfDWcomparison.pdfDARMsuppression.pdfControlLoop.pdf
                      Reply  Sat May 12 10:02:03 2018, rana, Update, General, More refinement of DARM control signal projection 
                         Reply  Sun May 20 17:43:01 2018, rana, Update, Electronics, How to choose resistors 
                      Reply  Mon May 14 18:58:32 2018, Kevin, Update, PonderSqueeze, Squeezing with no SRM homodyne_heatmap.pdfdisplacement_noise.pdfnoise_budget.pdf
                         Reply  Tue Oct 2 23:57:16 2018, gautam, Update, PonderSqueeze, Squeezing scenarios PonderSqueeze_NB_LORIN.pdf
                   Reply  Tue May 15 21:56:57 2018, gautam, Update, General, Stack measurement setup decommissioned 
                      Reply  Thu May 17 09:14:38 2018, Steve, Update, General, Stack measurement setup decommissioned Load_Cell_Measurement_Set_Up.jpgETMX_stack_up_down.png
Message ID: 13833     Entry time: Fri May 11 13:58:42 2018     In reply to: 13831     Reply to this: 13835   13846
Author: rana 
Type: Update 
Category: General 
Subject: More refinement of DARM control signal projection 

I think "OLG" trace is not labeled right; it would be good to see the actual OLG in addition to whatever that trace actually is.

Based on the first plot, however, my conclusion is that:

  1. we don't need the passive isolator to reduce the control signal; the control signal is dominated by f < 10 Hz.
  2. we should still look into isolators for the reduction of the f > 50 Hz stuff, just to make the overall DARM sensitivity better. But this does not have to be pneumatic since we no longer need 10 Hz isolation. It can instead be a solid piece of rubber to give us a ~20-30 Hz resonance. That would still give us a factor of 5-10 improvement above 100 Hz.
  3. In this case, we only need a mass estimate of the end chamber contents with an accuracy of ~25%. If we think we have that already, we don't need to keep doing the jacks-strain gauge adventure.