40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Sep 1 15:46:12 2017, Kira, Update, PEM, temp sensor update IMG_20170901_144729.jpg
    Reply  Tue Sep 5 17:18:17 2017, Kira, Update, PEM, temp sensor update IMG_20170905_144924.jpgIMG_20170905_165042.jpg
       Reply  Tue Sep 5 17:52:06 2017, Kira, Update, PEM, temp sensor update 
          Reply  Fri Sep 8 10:22:30 2017, Steve, Update, PEM, temp sensor update 
       Reply  Tue Sep 12 11:44:16 2017, Kira, Update, PEM, temp sensor update 
Message ID: 13303     Entry time: Fri Sep 8 10:22:30 2017     In reply to: 13296
Author: Steve 
Type: Update 
Category: PEM 
Subject: temp sensor update 

The weight of SS can with copper liner  is 12.2 kg

Is 1 Amp for the heating jacket going to be enough? We should have some headroom.

Quote:

to get the sensors to read the same values they have to be in direct thermal contact with the metal block - there can't be any adapter board in-between

for the 2nd attempt, I also recommend encasing it in a metal block rather than just one side. You can drill some 7-10 mm diameter holes in an aluminum or copper block. Then put the sensors in there and plug it up with some thermal paste.

 

ELOG V3.1.3-