Message ID: 13296     Entry time: Tue Sep 5 17:52:06 2017     In reply to: 13295     Reply to this: 13303
Author: Kira 
Type: Update 
Category: PEM 
Subject: temp sensor update 

to get the sensors to read the same values they have to be in direct thermal contact with the metal block - there can't be any adapter board in-between

for the 2nd attempt, I also recommend encasing it in a metal block rather than just one side. You can drill some 7-10 mm diameter holes in an aluminum or copper block. Then put the sensors in there and plug it up with some thermal paste.