40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Thu Aug 24 00:39:47 2017, gautam, Update, LSC, DRMI locking attempt
|
Fri Aug 25 01:20:52 2017, gautam, Update, LSC, DRMI locking attempt
|
Wed Aug 30 19:49:33 2017, gautam, Update, LSC, REFL55 demod board debugging
|
Thu Aug 31 00:52:52 2017, gautam, Update, LSC, REFL55 whitening board debugging 
|
Thu Aug 31 03:31:15 2017, gautam, Update, LSC, DRMI re-locked! 
|
Mon Sep 4 16:30:06 2017, gautam, Update, LSC, Oplev loop tweaking
|
Mon Sep 4 18:18:29 2017, rana, Update, LSC, dewhite switching: FOTON settings
|
Tue Sep 5 02:07:49 2017, gautam, Update, LSC, Low Noise DRMI attempt 
|
Wed Sep 6 23:06:30 2017, gautam, Update, LSC, Coil de-whitening switching investigation 
|
Fri Sep 15 16:00:33 2017, gautam, Update, LSC, Sensing measurement
|
Fri Sep 15 17:08:58 2017, gautam, Update, LSC, Coil de-whitening switching investigation
|
Sat Sep 16 10:56:19 2017, rana, Update, LSC, Coil de-whitening switching investigation
|
Fri Sep 22 18:12:27 2017, gautam, Update, LSC, DAC noise measurement (again)
|
Wed Sep 27 22:25:21 2017, gautam, Update, LSC, DAC noise measurement (again)
|
Thu Sep 28 06:35:44 2017, Chris, Update, LSC, DAC noise measurement (again)
|
Thu Sep 28 23:50:04 2017, gautam, Update, LSC, DAC noise measurement (again)
|
Fri Sep 29 18:36:25 2017, gautam, Update, LSC, DAC noise measurement (again)
|
Tue Oct 3 01:32:39 2017, gautam, Update, LSC, Laser intensity noise coupling to MICH (simulated) 
|
|
Message ID: 13290
Entry time: Mon Sep 4 18:18:29 2017
In reply to: 13289
|
Author: |
rana |
Type: |
Update |
Category: |
LSC |
Subject: |
dewhite switching: FOTON settings |
|
|
not immediately necessary, since you have already got it sort of working, but one of these days we should optimize this for real. In the past, we used to do this by putting a o'scope on the coil Vmon during the switching to catch the transient w/ triggering. We download the data/picture via ethernet. Run for loop on tolerance to see what's what.
- Went into the Foton filter banks for all the coil output filters, and modified the "Output" settings to be on "Input crossing", with a "Tolerance" of 10 and a "Timeout" of 3 seconds. These settings are to facilitate smooth transition between the two signal paths (without and with coil-dewhitening). The parameters chosen were arbitrary and not optimized in any systematic manner.
|
|