Entry  Sat Aug 5 11:04:07 2017, gautam, Update, SUS, MC1 glitches return MC1_glitch_Aug42017.pngMC1_glitch_zoomed.png
    Reply  Wed Aug 9 15:15:47 2017, gautam, Update, SUS, MC1 glitches return MC1_glitch.png
       Reply  Thu Aug 10 21:01:43 2017, gautam, Update, SUS, MC1 glitches debugging MC_QUAD_10AUG2017.jpgMCR_10AUG2017.jpg
          Reply  Fri Aug 11 12:32:46 2017, gautam, Update, SUS, MC1 glitches debugging MC1_freeswinging.pdfMC1_inmatrix.pngMC1_sensors.png
             Reply  Fri Aug 11 17:36:47 2017, gautam, Update, SUS, MC1 <--> MC3 MC1_glitch_watchdog_shutdown.png
                Reply  Sat Aug 12 14:09:36 2017, gautam, Update, SUS, Glitches stay on MC1 MC1_glitch.pngMC_12hr_trend.pngMC1_glitches_intermittent.png
                   Reply  Sat Aug 12 22:35:22 2017, rana, Update, SUS, Glitches stay on MC1 
                      Reply  Sun Aug 13 01:35:09 2017, Koji, Update, SUS, Glitches stay on MC1 
                         Reply  Mon Aug 14 20:01:38 2017, gautam, Update, SUS, Glitches stay on MC1 
                Reply  Wed Aug 16 19:50:17 2017, gautam, Update, SUS, MC1 <--> MC3 switched back 
                   Reply  Thu Aug 17 11:17:49 2017, gautam, Update, SUS, MC1 <--> MC3 switched back MC1_misaligned.pngMC1_glitch.png
                      Reply  Thu Aug 17 17:33:01 2017, gautam, Update, SUS, MC1 <--> MC3 switched back 
Message ID: 13200     Entry time: Sat Aug 12 22:35:22 2017     In reply to: 13199     Reply to this: 13201
Author: rana 
Type: Update 
Category: SUS 
Subject: Glitches stay on MC1 

To add to Gautam's entry: we swapped the cables at the coil driver side (these are the ones that go from coil driver to sat box). In this state, damping is not useable since the MC1 servos would drive MC3.

~70 counts in the sensor means ~70 microns of motion. Since the watchdogs are off and the coil drivers are swapped, this can't be laser beam getting in to the sensors.

WE have to consider that these are some real strain release type events happening in the suspension wire or wire standoff, so may require a vent to inspect and possible repair MC1.