40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Sat Aug 5 11:04:07 2017, gautam, Update, SUS, MC1 glitches return MC1_glitch_Aug42017.pngMC1_glitch_zoomed.png
    Reply  Wed Aug 9 15:15:47 2017, gautam, Update, SUS, MC1 glitches return MC1_glitch.png
       Reply  Thu Aug 10 21:01:43 2017, gautam, Update, SUS, MC1 glitches debugging MC_QUAD_10AUG2017.jpgMCR_10AUG2017.jpg
          Reply  Fri Aug 11 12:32:46 2017, gautam, Update, SUS, MC1 glitches debugging MC1_freeswinging.pdfMC1_inmatrix.pngMC1_sensors.png
             Reply  Fri Aug 11 17:36:47 2017, gautam, Update, SUS, MC1 <--> MC3 MC1_glitch_watchdog_shutdown.png
                Reply  Sat Aug 12 14:09:36 2017, gautam, Update, SUS, Glitches stay on MC1 MC1_glitch.pngMC_12hr_trend.pngMC1_glitches_intermittent.png
                   Reply  Sat Aug 12 22:35:22 2017, rana, Update, SUS, Glitches stay on MC1 
                      Reply  Sun Aug 13 01:35:09 2017, Koji, Update, SUS, Glitches stay on MC1 
                         Reply  Mon Aug 14 20:01:38 2017, gautam, Update, SUS, Glitches stay on MC1 
                Reply  Wed Aug 16 19:50:17 2017, gautam, Update, SUS, MC1 <--> MC3 switched back 
                   Reply  Thu Aug 17 11:17:49 2017, gautam, Update, SUS, MC1 <--> MC3 switched back MC1_misaligned.pngMC1_glitch.png
                      Reply  Thu Aug 17 17:33:01 2017, gautam, Update, SUS, MC1 <--> MC3 switched back 
Message ID: 13178     Entry time: Wed Aug 9 15:15:47 2017     In reply to: 13168     Reply to this: 13187
Author: gautam 
Type: Update 
Category: SUS 
Subject: MC1 glitches return 

Happened again just now, although the characteristics of the glitch are very different from the previous post, its less abrupt. Only actuation on MC1 at this point was local damping.

Attachment 1: MC1_glitch.png  38 kB  | Hide | Hide all
MC1_glitch.png
ELOG V3.1.3-