40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Apr 26 01:00:23 2017, gautam, Update, General, DRMI locking  DRMI_1f_26April2017.pngIMG_7373.JPGIMG_7372.JPG
    Reply  Fri Apr 28 02:04:36 2017, gautam, Update, General, DRMI locking  20170428_DRMI.pdf
       Reply  Fri Apr 28 13:56:26 2017, rana, Update, General, DRMI locking  
          Reply  Mon May 1 16:29:51 2017, gautam, Update, General, DRMI locking  LSC_sensingNoise.pdfIMG_7378.JPG
             Reply  Mon May 1 21:45:54 2017, ericq, Update, General, DRMI locking  
                Reply  Wed May 3 18:45:45 2017, rana, Update, General, DRMI locking  
                   Reply  Thu May 4 19:03:15 2017, gautam, Update, General, DRMI locking - preliminary MICH NB DRMI_noArms_April30.pdfMICH_OLTF.pdfC1NB_disp_40m_MICH_NB_30_April_2017.pdf40m_REFL_RFPDs_efficiency.pdf
                      Reply  Fri May 5 10:13:02 2017, ericq, Update, General, MICH NB questions 
                         Reply  Fri May 5 12:10:53 2017, gautam, Update, General, MICH NB questions C1NB_disp_40m_MICH_NB_30_April_2017.pdf
                            Reply  Sat May 6 21:52:11 2017, rana, Update, General, MICH NB questions 
                               Reply  Wed May 10 01:56:06 2017, gautam, Update, General, MICH NB - OL coupling OL_BS_ITM_comp.pdfC1NB_disp_40m_MICH_NB_8_May_2017.pdfC1NB_disp_40m_MICH_NB_10_May_2017.pdf
                                  Reply  Wed May 10 16:53:38 2017, rana, Update, General, MICH NB - OL coupling 
                                  Reply  Wed May 10 17:17:05 2017, gautam, Update, General, DAC / Coil Driver noise actuation.jpg
                                     Reply  Wed May 10 17:46:44 2017, gautam, Update, General, DAC / Coil Driver noise - SRM coil driver + dewhite board removed 
                                        Reply  Thu May 11 09:45:46 2017, rana, Update, General, DAC / Coil Driver noise - SRM coil driver + dewhite board removed 
                                           Reply  Thu May 11 18:59:22 2017, gautam, Update, General, SRM coil driver + dewhite board initial survey D010001-B_40m.pdfD000183-C8_40m.pdf
                                              Reply  Fri May 12 01:36:04 2017, gautam, Update, General, SRM coil driver + dewhite board LISO modeling SRM_bypass_plus_CoilDriver.pdfliso.zip
                                                 Reply  Fri May 12 12:34:55 2017, gautam, Update, General, ITM and BS coil driver + dewhite board pulled out 
                                                    Reply  Fri May 12 18:50:08 2017, gautam, Update, General, ITM and BS coil driver + dewhite board pulled out 
                                                       Reply  Mon May 22 13:37:01 2017, gautam, Update, General, DAC noise estimate DAC_noise_model.pdfC1NB_disp_40m_MICH_NB_22_May_2017.pdf
                                                          Reply  Tue May 23 22:58:23 2017, gautam, Update, General, De-Whitening board noises ITMX_deWhite_ch1_noise.pdf
                                                             Reply  Thu May 25 19:27:29 2017, gautam, Update, General, Coil driver board noises  coilDriverNoises.pdf
                                                                Reply  Mon May 29 16:47:38 2017, gautam, Update, General, Coil driver boards reinstalled deWhitening_consolidated.pdfdeWhitening_measurements.zipETMY_OL.png
                                                                   Reply  Tue May 30 16:02:59 2017, gautam, Update, General, Coil driver boards reinstalled OL_noiseInjection.pdf
                                                                      Reply  Thu Jun 1 00:10:15 2017, gautam, Update, General, Coil driver boards reinstalled 
       Reply  Fri Apr 28 19:32:06 2017, gautam, Update, General, DRMI locking - PRCL angular FF PRC_angularFF.pdfPRC_TFs.pdf
Message ID: 12975     Entry time: Fri May 5 12:10:53 2017     In reply to: 12974     Reply to this: 12976
Author: gautam 
Type: Update 
Category: General 
Subject: MICH NB questions 

Quote:
Is suspension thermal noise missing? I take it "Thermal" refers just to thermal things going on in the optic, since I don't see any peaks at the bounce/roll modes as I would expect from suspension thermal noise. What goes into the GWINC calculation of seismic noise? Does it include real 40m ground motion data and our seismic stacks? I'm surprised to see such a sharp corner in the "Dark Noise" trace, did you apply the OLG correction to a measured dark noise ASD? (The OLG correction only needs to be applied to the in-lock error signals to recover open loop behavior, there is no closed loop when you're measuring the dark noise so nothing to correct for.)


I've included the suspension thermal noise in the "Thermal" trace, but I guess the GWINC file I've been using to generate this trace only computes the thermal noise for the displacement DoF. I think this paper has the formulas to account for them, I will look into including these.

For the seismic noise, I've just been using the seis40.mat file from the 40m SVN. I think it includes a model of our stacks, but I did not re-calculate anything with current seismometer spectra. In the NB I updated yesterday, however, I think I was off by a factor of sqrt(3) as I had only included the seismic noise from 1 suspended optic. I've corrected this in the attached plot.

For the dark noise, you are right, I had it grouped in the wrong dictionary in the code so it was applying the OLG inversion. I've fixed this in the attached plot.
Attachment 1: C1NB_disp_40m_MICH_NB_30_April_2017.pdf  772 kB  Uploaded Fri May 5 19:30:09 2017  | Show | Show all
ELOG V3.1.3-