40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Wed Apr 26 01:00:23 2017, gautam, Update, General, DRMI locking  
|
Fri Apr 28 02:04:36 2017, gautam, Update, General, DRMI locking
|
Fri Apr 28 13:56:26 2017, rana, Update, General, DRMI locking
|
Mon May 1 16:29:51 2017, gautam, Update, General, DRMI locking 
|
Mon May 1 21:45:54 2017, ericq, Update, General, DRMI locking
|
Wed May 3 18:45:45 2017, rana, Update, General, DRMI locking
|
Thu May 4 19:03:15 2017, gautam, Update, General, DRMI locking - preliminary MICH NB   
|
Fri May 5 10:13:02 2017, ericq, Update, General, MICH NB questions
|
Fri May 5 12:10:53 2017, gautam, Update, General, MICH NB questions
|
Sat May 6 21:52:11 2017, rana, Update, General, MICH NB questions
|
Wed May 10 01:56:06 2017, gautam, Update, General, MICH NB - OL coupling  
|
Wed May 10 16:53:38 2017, rana, Update, General, MICH NB - OL coupling
|
Wed May 10 17:17:05 2017, gautam, Update, General, DAC / Coil Driver noise
|
Wed May 10 17:46:44 2017, gautam, Update, General, DAC / Coil Driver noise - SRM coil driver + dewhite board removed
|
Thu May 11 09:45:46 2017, rana, Update, General, DAC / Coil Driver noise - SRM coil driver + dewhite board removed
|
Thu May 11 18:59:22 2017, gautam, Update, General, SRM coil driver + dewhite board initial survey 
|
Fri May 12 01:36:04 2017, gautam, Update, General, SRM coil driver + dewhite board LISO modeling 
|
Fri May 12 12:34:55 2017, gautam, Update, General, ITM and BS coil driver + dewhite board pulled out
|
Fri May 12 18:50:08 2017, gautam, Update, General, ITM and BS coil driver + dewhite board pulled out
|
Mon May 22 13:37:01 2017, gautam, Update, General, DAC noise estimate 
|
Tue May 23 22:58:23 2017, gautam, Update, General, De-Whitening board noises
|
Thu May 25 19:27:29 2017, gautam, Update, General, Coil driver board noises
|
Mon May 29 16:47:38 2017, gautam, Update, General, Coil driver boards reinstalled  
|
Tue May 30 16:02:59 2017, gautam, Update, General, Coil driver boards reinstalled
|
Thu Jun 1 00:10:15 2017, gautam, Update, General, Coil driver boards reinstalled
|
Fri Apr 28 19:32:06 2017, gautam, Update, General, DRMI locking - PRCL angular FF 
|
|
Message ID: 12955
Entry time: Fri Apr 28 13:56:26 2017
In reply to: 12954
Reply to this: 12960
|
Author: |
rana |
Type: |
Update |
Category: |
General |
Subject: |
DRMI locking |
|
|
one of these signals does not look like the others: explanation? |