Thu Feb 5 21:42:28 2009, Yoichi, Update, PSL, My thoughts on ISS
|
Fri Feb 6 09:52:35 2009, Kakeru, Update, PSL, Current shunt transfar function 
|
Fri Feb 6 23:23:48 2009, Kakeru, Yoichi, Update, PSL, ISS is fixed 
|
Mon Feb 9 16:02:42 2009, Yoichi, Update, PSL, PSL relative intensity noise 
|
Tue Feb 10 23:36:25 2009, Kakeru, Update, PSL, PA current and laser output
|
Wed Feb 11 07:28:25 2009, Yoichi, Update, PSL, PA current and laser output
|
Wed Feb 11 23:51:53 2009, Kakeru, Update, PSL, PA current and laser output
|
|
Message ID: 1295
Entry time: Wed Feb 11 23:51:53 2009
In reply to: 1291
|
Author: |
Kakeru |
Type: |
Update |
Category: |
PSL |
Subject: |
PA current and laser output |
|
|
I attached a plot of ISS monitor PD and MOPA output to PA current.
The both end of PA current (26.0353[A] and 28.4144[A]) correspond to the slider value of -2.0 and 1.0 .
It looks that we must use MOPA with PA current below 27.5[A]. |
|
|