40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Mon Mar 27 20:41:21 2017, rana, gautam, Update, PSL, PMC DAQ assay for feed-forward integration  
|
Wed Mar 29 11:41:19 2017, gautam, Update, PSL, PMC DAQ assay for feed-forward integration  
|
Thu Mar 30 00:16:09 2017, gautam, Update, PSL, PSL NPRO PZT calibration
|
Mon Apr 3 17:25:13 2017, gautam, Update, PSL, PSL NPRO PZT calibration 
|
Tue Apr 4 17:27:58 2017, rana, Update, PSL, PSL NPRO PZT calibration
|
Thu Mar 30 18:11:01 2017, rana, Update, PSL, PMC DAQ assay for feed-forward integration
|
Mon Apr 3 23:07:09 2017, gautam, Update, PSL, PMC DAQ assay for feed-forward integration
|
Tue Apr 11 00:38:37 2017, gautam, Update, PSL, PMC demod moved off servo board
|
Wed Apr 12 00:36:53 2017, gautam, Update, PSL, PMC demod moved off servo board 
|
Tue Apr 18 01:01:03 2017, gautam, Update, PSL, PMC OLTF measured, DAQ channels calibrated 
|
Tue Apr 18 16:10:00 2017, gautam, Update, PSL, PMC OLTF measured, DAQ channels calibrated
|
Tue Apr 18 23:37:15 2017, rana, Update, PSL, PMC OLTF measured, DAQ channels calibrated
|
Wed Apr 19 15:13:30 2017, gautam, Update, PSL, PMC/MCL multicoherence
|
|
Message ID: 12947
Entry time: Wed Apr 19 15:13:30 2017
In reply to: 12945
|
Author: |
gautam |
Type: |
Update |
Category: |
PSL |
Subject: |
PMC/MCL multicoherence |
|
|
I used a one hour stretch of data from last night to look at coherence between the PMC control signal and MCL, to see if the former can be used as a witness channel in some frequency band for MCL stabilization. Here is a plot of the predicted subtraction and coherence, made using EricQs pynoisesub code. I had thought about adopting the greedy channel ranking algorithm that Eric has been developing for noise subtraction in site data, but since I am just considering 3 witness channels, I figured this straight up comparison between different sets of witness channels was adequate. Looks like we get some additional coherence with MCL by adding the PMC control signal to the list of witness channels, there is about a factor of a few improvement in in the 1-2Hz band...

|
|
|