Thu Feb 16 13:29:38 2017, gautam, Summary, General, Alternative Calibration Scheme
|
Thu Feb 16 21:55:47 2017, rana, Summary, General, Alternative Calibration Scheme
|
Tue Feb 21 13:51:35 2017, Craig, Summary, General, Alternative Calibration Scheme
|
Wed Feb 22 10:16:54 2017, rana, Summary, General, Alternative Calibration Scheme
|
|
Message ID: 12845
Entry time: Wed Feb 22 10:16:54 2017
In reply to: 12842
|
Author: |
rana |
Type: |
Summary |
Category: |
General |
Subject: |
Alternative Calibration Scheme |
|
|
OK, but the questions still stands: "Assuming we want a 1% calibration at 50-500 Hz, what is the requirement on the frequency noise PSD curve?"
Quote: |
We get SNR in two ways: the amplitude of applied force and the integration time. So we are limited in two ways: stability of the lock to applied forces and time of locklosses / calibration fluctuations.
|
|