Wed Jan 25 15:41:29 2017, Lydia, Update, IMC, 29.5 MHz modulation depth measurement plan
|
Wed Jan 25 17:30:03 2017, Koji, Update, IMC, 29.5 MHz modulation depth measurement plan
|
Wed Jan 25 19:39:07 2017, gautam , Update, IMC, 29.5 MHz modulation depth measurement plan
|
Fri Jan 27 21:25:11 2017, Lydia, Update, IMC, 29.5 MHz modulation depth
|
Sat Jan 28 01:25:51 2017, gautam, Update, IMC, 29.5 MHz modulation depth 
|
Mon Jan 30 19:07:48 2017, gautam, Update, IMC, RF AM stabilization box pulled out
|
Tue Jan 31 01:07:20 2017, Lydia, Update, IMC, RF AM stabilization box pulled out
|
Tue Jan 31 13:46:34 2017, rana, Update, IMC, RF AM stabilization box pulled out
|
Tue Jan 31 14:17:48 2017, gautam, Update, IMC, RF AM stabilization box pulled out
|
Tue Jan 31 22:07:13 2017, gautam, Update, IMC, RF AM stabilization box revamp
|
Tue Jan 31 22:28:39 2017, Lydia, Update, IMC, RF AM stabilization box pulled out
|
Wed Feb 1 11:51:19 2017, Koji, Update, IMC, RF AM stabilization box pulled out
|
Wed Feb 1 16:45:56 2017, Lydia, Update, IMC, RF AM stabilizer box Modification Plan
|
Wed Feb 1 20:49:34 2017, rana, Update, IMC, RF AM stabilizer box Modification Plan
|
Wed Feb 1 23:13:30 2017, Lydia, Update, IMC, RF AM stabilizer box Modification Plan
|
Sun Feb 5 21:56:50 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement
|
Tue Feb 7 12:01:10 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement
|
Tue Feb 7 17:00:55 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement
|
Fri Feb 10 02:14:10 2017, gautam, Update, IMC, 29.5 MHz stabilizer box replacement
|
Fri Feb 10 11:41:43 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement
|
Fri Feb 10 13:24:28 2017, rana, Update, IMC, 29.5 MHz stabilizer box replacement
|
Fri Feb 10 18:21:21 2017, gautam, Update, IMC, IMC Demod board
|
Fri Feb 10 19:32:15 2017, Koji, Update, IMC, IMC Demod board
|
Sun Feb 12 01:16:57 2017, gautam, Update, IMC, IMC length loop - summary of changes    
|
Mon Feb 13 11:55:14 2017, rana, Update, IMC, IMC length loop - summary of changes
|
Mon Feb 13 13:34:44 2017, gautam, Update, IMC, IMC length loop - bad SMA cable replaced
|
Mon Feb 13 19:44:55 2017, Lydia, Update, IMC, Front panel for 29.5 MHz amplifier box
|
Wed Mar 1 21:15:40 2017, Lydia, Update, IMC, Front panel for 29.5 MHz amplifier box
|
Wed Mar 1 23:56:09 2017, gautam, Update, IMC, Front panel for 29.5 MHz amplifier box
|
Thu Mar 2 20:32:18 2017, Lydia, Update, IMC, Front panel for 29.5 MHz amplifier box
|
Wed Feb 15 23:54:13 2017, gautam, Update, IMC, IMC saga continues... 
|
Fri Feb 17 20:10:18 2017, gautam, Update, IMC, WFS servos turned back on
|
Sat Feb 18 14:09:06 2017, rana, Update, IMC, WFS servos turned back on
|
Sat Feb 18 21:50:48 2017, gautam, Update, IMC, WFS servos turned back on
|
Sun Mar 5 12:41:23 2017, gautam, Update, IMC, WFS servo-steppin  
|
|
Message ID: 12784
Entry time: Wed Feb 1 16:45:56 2017
In reply to: 12783
Reply to this: 12785
|
Author: |
Lydia |
Type: |
Update |
Category: |
IMC |
Subject: |
RF AM stabilizer box Modification Plan |
|
|
Here's what I'm planning to do to the RF AM stabilizer box. I'm going to take out several of the components along the path to the EOM (comments in green), including the dead ERA-4 and ERA-5 amplifiers, the variable attenuator which is controlled by a switch that can't be accessed outside the box, and the feedback path from the daughter board servo. I'm arranging things so that the output of the HELA-10 does not exceed the maximum output power.
I wasn't quite as sure what to do about the path to the ASC box (comments in blue). I talked with Gautam and he said this gets split equally between several singals, one of which goes to the LO of the demod board which expects -10 dBm and currently gets -12 dBm (can go up to -8 by turning switch). So maybe we don't actually want the signal to be anywhere near +27 dBm at the output. The plans for the box are here, it looks like +27 in will end up with +10 at each output, which is way more than what's currently coming out. But maybe this needs to be increased to match the other path?
Also we haven't measured the actual response of the variable attenuator U4 for various switch positions; it's the same model as the one I'm removing from the EOM path and that one had slightly different behavior for different switch positions than what the spec sheet says. Same goes for the HELA-10 units along this path: what is their actual gain? So perhaps these should be measured and then a single attenuator should be chosen to get the right output signal level. Alternatively it could just be left alone, if it is at an OK level right now. Advice on what to do here would be appreciated.
I'll work on the EOM path tonight and wait for feedback on the rest of it.
EDIT: Gautam pointed out that there's some insertion loss from the components I'll be removing that hasn't been accounted for. Also the plans have been updated to reflect that I'm replacing AT5 with a 1dB attenuator (from 6 dB). |
|
|