Mon Jan 23 15:16:52 2017, gautam, Update, Optical Levers, ETMY Oplev HeNe needs to be replaced
|
Mon Jan 23 17:24:26 2017, Steve, Update, Optical Levers, ETMY Oplev HeNe is running hot
|
Tue Jan 24 17:52:15 2017, Craig, Update, Optical Levers, ETMY Oplev HeNe is replaced 6x
|
Wed Mar 1 08:28:04 2017, Steve, Update, Optical Levers, oplev laser RIN 
|
Thu Mar 2 13:59:04 2017, Steve, Update, Optical Levers, oplev laser RIN
|
Thu Mar 2 17:58:45 2017, rana, Update, Optical Levers, oplev laser RIN
|
Fri Mar 3 17:24:21 2017, Steve, Update, Optical Levers, oplev laser RIN
|
Wed Mar 8 15:28:37 2017, Steve, Update, Optical Levers, oplev laser RIN 
|
Mon Apr 10 09:49:32 2017, Steve, Update, Optical Levers, oplev laser RIN test planning
|
Mon Apr 10 14:21:57 2017, rana, Update, Optical Levers, oplev laser RIN test planning
|
Tue May 30 13:36:58 2017, Steve, Update, Optical Levers, ETMY Oplev HeNe is replaced
|
Tue Jan 15 18:16:00 2019, gautam, Update, Optical Levers, ETMY Oplev HeNe needs replacement 
|
Fri Jan 18 12:52:07 2019, gautam, Update, Optical Levers, BS/PRM Oplev HeNe replaced
|
|
Message ID: 12746
Entry time: Mon Jan 23 15:16:52 2017
Reply to this: 12747
|
Author: |
gautam |
Type: |
Update |
Category: |
Optical Levers |
Subject: |
ETMY Oplev HeNe needs to be replaced |
|
|
On the control room monitors, I noticed that the IR TEM00 spot was moving around rather a lot in the Y arm. The last time this happened had something to do with the ETMY Oplev, so I took a look at the 30 day trend of the QPD sum, and saw that it was decaying steeply (Steve will update with a long term trend plot shortly). I noticed the RIN also seemed rather high, judging by how much the EPICS channel reading for the QPD sum was jumping around. Attached are the RIN spectra, taken with the OL spot well centered on the QPD and the arms locked to IR. Steve will swap the laser out if it is indeed the cluprit. |
|
|