40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Mon Jan 16 23:25:07 2017, gautam, Update, SUS, MC1 SUS electronics investigation  
|
Tue Jan 17 21:29:52 2017, gautam, Update, SUS, MC1 SUS electronics investigation
|
Wed Jan 18 11:40:54 2017, gautam, Update, SUS, MC1 SUS electronics investigation  
|
Wed Jan 18 14:23:47 2017, gautam, Update, SUS, MC1 SUS electronics investigation
|
Wed Jan 18 18:44:53 2017, gautam, Update, SUS, MC1 SUS electronics investigation
|
Thu Jan 19 08:25:12 2017, Steve, Update, SUS, MC1 SUS electronics investigation 
|
Thu Jan 19 12:00:10 2017, gautam, Update, SUS, MC1 SUS electronics investigation
|
Thu Jan 19 19:56:09 2017, rana, Update, SUS, MC1 SUS electronics investigation
|
Fri Jan 20 11:16:30 2017, gautam, Update, SUS, MC1 SUS electronics investigation
|
Mon Jan 23 10:24:01 2017, Steve, Update, SUS, MC1 SUS electronics investigation
|
|
Message ID: 12728
Entry time: Tue Jan 17 21:29:52 2017
In reply to: 12725
Reply to this: 12731
|
Author: |
gautam |
Type: |
Update |
Category: |
SUS |
Subject: |
MC1 SUS electronics investigation |
|
|
Quote: |
After some investigation, Rana found that on the AA filter board end, one of the 4pin LEMOs from the whitening board had one of its wires come unstuck from where it was soldered (this presumably happened while we were squishing cables tonight, as the LR channel was fine before that). Also, there was no heat shrink used on any of the solder joints.
|
The faulty cable has been re-soldered (with heat shrink) and replaced. All 5 sensor signals appear normal on dataviewer now. I am leaving things in this state for the night, let us see if the glitches return overnight.
PSL shutter remains closed |