Message ID: 12551
Entry time: Tue Oct 11 13:30:49 2016
In reply to: 12549
Reply to this: 12552
|
Author: |
gautam |
Type: |
Update |
Category: |
SUS |
Subject: |
PRM LR problematic again |
|
|
Perhaps the problem is electrical? The attached plot shows a downward trend for the LR sensor output over the past 20 days that is not visible in any of the other 4 sensor signals. The Al foil was shorting the electrical contacts for nearly 2 months, so perhaps some part of the driver circuit needs to be replaced? If so a Satellite Box swap should tell us more, I will switch the PRM and SRM satellite boxes. It could also be a dying LED on the OSEM itself I suppose. If we are accessing the chamber, we should come up with a more robust insulating cap solution for the OSEMs rather than this hacky Al foil + kapton arrangement.
The PRM and SRM Satellite boxes have been switched for the time being. I had to adjust some of the damping loop gains for both PRM and SRM and also the PRM input matrix to achieve stable damping as the PRM Satellite box has a Side sensor which reads out 0-10V as opposed to the 0-2V that is usually the case. Furthermore, the output of the LR sensor going into the input matrix has been turned off.

|