[Gautam, Johannes]
I scripted a series of YARM DC reflectivity measurements last night alternating between locked state and unlocked state (with ETMY misaligned) for measuring the after-vent armloss. The general procedure is based on elog 11810, but I'll also give a brief summary here.
- To measure the locked reflectivity the dither script is executed with a stop condition that depends on the rms values of its error signals.
- The dithering is stopped, and while the arm is locked the reflected power is recorded from both POX/POY DC and ASDC, as well as the mode cleaner transmission for normalization.
- The arm locking is switched off, and ETMY moved to is 'misaligned' position. This gets rid off unwanted mode flashes.
- In the unlocked state the same quantities are recorded.
- Rinse and repeat for a set number of times (for this run I set it to 100 and left the interferometer alone).
I did this back in June (but strangely never posted what I found, shame on me). What I found back then was a YARM loss of 237 ppm +/- 41 ppm and an XARM loss of 501 ppm +/- 105 ppm
Last night's data indicates a YARM loss of 143 ppm +/- 24 ppm after cleaning with first contact.  