Fri Sep 16 19:48:52 2016, Lydia, Update, General, Alignment status
|
Sat Sep 17 02:00:23 2016, rana, Update, SUS, Alignment status
|
Sat Sep 17 16:51:01 2016, Lydia, Update, SUS, Alignment status
|
Sun Sep 18 16:18:05 2016, rana, Update, SUS, Alignment status
|
Mon Sep 19 11:11:43 2016, ericq, Update, SUS, Alignment status
|
Mon Sep 19 22:03:10 2016, ericq, Update, General, Further recovery progress
|
Tue Sep 20 10:45:06 2016, rana, Update, General, Further recovery progress
|
Wed Sep 21 01:08:02 2016, ericq, Update, General, Further recovery progress
|
Wed Sep 21 09:04:57 2016, Steve, Update, General, 8 hours recovery progress
|
Thu Sep 22 15:44:21 2016, Steve, Update, General, ITMX magnets are stucked again
|
Thu Sep 22 20:01:47 2016, Lydia, Update, General, ITMX freed, all optics kicked
|
Thu Sep 22 22:52:08 2016, ericq, Update, General, Damping found to be on
|
|
Message ID: 12501
Entry time: Sat Sep 17 02:00:23 2016
In reply to: 12500
Reply to this: 12502
|
Author: |
rana |
Type: |
Update |
Category: |
SUS |
Subject: |
Alignment status |
|
|
All is not lost. I've stuck and unstuck optics around a half dozen times. Can you please post the zoomed in time series (not trend) from around the time it got stuck? Sometimes the bias sliders have to be toggles to make the bias correct. From the OSEM trend it seems like it got a large Yaw bias. May also try to reseat the satellite box cables and the cable from the coil driver to the cable breakout board in the back of the rack. |