40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Thu Jun 19 17:11:52 2014, steve, Update, PEM, it is warmer today
|
Mon Sep 22 10:18:58 2014, steve, Update, PEM, lab temperatures
|
Tue Nov 4 10:40:46 2014, steve, Update, PEM, lab temperatures and particle counts
|
Mon Dec 8 09:03:49 2014, steve, Update, PEM, air cond maintenance tomorrow morning
|
Wed Dec 10 14:26:13 2014, steve, Update, PEM, air cond maintenance and particle plot
|
Fri Sep 16 14:15:28 2016, steve, Update, PEM, partical counts
|
Fri Dec 2 10:57:08 2016, steve, Update, PEM, door seal replaced
|
Fri Feb 17 10:56:12 2017, steve, Update, PEM, particle counter moved into PSL enclousure  
|
Tue Feb 21 10:08:35 2017, steve, Update, PEM, noisy morning
|
Wed Feb 22 08:54:17 2017, steve, Update, PEM, PSL enclousure particle count
|
Tue Feb 28 08:04:48 2017, steve, Update, PEM, ETMX damping recovered
|
Tue Feb 28 18:25:22 2017, rana, Update, PEM, ETMX damping recovered
|
Thu Mar 16 08:22:16 2017, Steve, Update, SUS, ETMX damping
|
Mon Aug 28 10:51:21 2017, steve, Update, SUS, ETMX damping recovered
|
Mon Mar 20 17:12:08 2017, steve, Update, PEM, particle counter inside of PSL enclousure
|
Mon Mar 27 15:36:27 2017, Steve, Update, PEM, particle counter inside of PSL enclousure
|
|
Message ID: 12498
Entry time: Fri Sep 16 14:15:28 2016
In reply to: 10773
Reply to this: 12656
12836
|
Author: |
steve |
Type: |
Update |
Category: |
PEM |
Subject: |
partical counts |
|
|
South end flow bench and both clean room assembly flow benches measured zero counts for 0.3 and 0.5 micron size particales.
The counting efficiency of 0.5 micron is 100%
location
|
0.3 micron particles / cf min
|
0.5 micron particles / cf min
|
counter
MET
ONE
|
effect |
ITMY table |
35,000 |
3,000 |
#3 |
|
ETMY table |
35,000 |
3,000 |
#3 |
|
ITMX table |
24,000 |
2,400 |
#3 |
PSL HEPA at 50V |
ITMX wall mounted |
- |
2,500 |
#1 |
|
ETMX work bench |
5,000 |
600 |
#3 |
flow bench on |
The PSL HEPA performance was measured at the center of the table with MET ONE #3
Voltage Variac |
particles /cf min |
particles /cf min |
particle size |
0.3 micron |
0.5 micron |
60 |
0 |
0 |
50 |
10 |
0 |
40 |
1,400 |
110 |
20 |
1,400 |
100 |
|
|
|