Thu Aug 25 15:43:29 2016, ericq, Update, SUS, OSEM issues 
|
Thu Aug 25 20:07:35 2016, gautam, Update, SUS, OSEM issues - maybe resolved?
|
Tue Sep 6 18:21:13 2016, rana, Update, SUS, OSEM issues
|
Tue Sep 6 20:52:42 2016, gautam, Update, SUS, OSEM issues
|
Wed Sep 7 17:23:26 2016, gautam, Update, SUS, PRM LR fixed for now 
|
Tue Oct 11 00:43:50 2016, ericq, Update, SUS, PRM LR problematic again  
|
Tue Oct 11 10:15:04 2016, Steve, Update, SUS, PRM LR problematic again
|
Tue Oct 11 13:30:49 2016, gautam, Update, SUS, PRM LR problematic again
|
Wed Oct 12 13:34:28 2016, gautam, Update, SUS, PRM LR problematic again
|
|
Message ID: 12475
Entry time: Tue Sep 6 20:52:42 2016
In reply to: 12472
Reply to this: 12476
|
Author: |
gautam |
Type: |
Update |
Category: |
SUS |
Subject: |
OSEM issues |
|
|
The modes look like they're at the right frequencies, so pointing more and more towards a LED or satellite box issue.
|
We peeked into the BS-PRM chamber via the ITMX chamber to see if we could shed any light on this situation. It's hard to get a picture that is in focus, but it looks quite clear that the LR LED (in the lower left when viewed from the HR side) isn't anywhere near as bright as the rest (see Attachment #1). Various hypothesis include failed LED / piece of Al foil blocking the LED / teflon aperture slipped over the LED. But looks like we can't solve this without opening up the BS-PRM chamber. The plan tomorrow is to open up the chamber, pull out the problematic coil. Once we have a better idea of what is going wrong, we can decide what the appropriate course of action is - replace the OSEM or something else.
As part of the diagnosis, I switched the PRM and SRM satellite boxes earlier today evening around 6pm. They remain in this switched state for now.
Steve, we plan to take the BS-PRM heavy door off tomorrow morning. |
|
|