Message ID: 12472     Entry time: Tue Sep 6 18:21:13 2016     In reply to: 12441     Reply to this: 12475
Author: rana 
Type: Update 
Category: SUS 
Subject: OSEM issues 

I looked at the PRM free swing spectra. The modes look like they're at the right frequencies, so pointing more and more towards a LED or satellite box issue.

  7/2011 9/2016
POS 0.993 0.94
PIT 0.612 0.60
YAW 0.833 0.76
SIDE 0.999 0.993

Some of the frequencies have changed between the 2011 in-vac measurement and our 2016 in-air measurement, but that seems within usual parameters.