40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Aug 3 15:36:57 2016, Steve, Update, SUS, Ni plated magnets & epoxy ordered 
    Reply  Fri Aug 5 09:38:12 2016, Steve, Update, SUS, SOS sus wire ordered 0.0017.jpg0.0017spec.jpg
       Reply  Mon Aug 15 14:29:32 2016, Steve, Update, SUS, SOS sus wire ordered 
          Reply  Mon Aug 15 14:34:33 2016, ericq, Update, SUS, SOS sus wire ordered 
          Reply  Wed Aug 17 22:09:04 2016, rana, Update, SUS, SOS sus wire ordered 
       Reply  Thu Aug 18 15:16:09 2016, Steve, Update, SUS, SOS sus wire is in 0.0017new.jpg
          Reply  Wed Nov 30 14:53:56 2016, Steve, Update, SUS, new sus wire stored in N2 filled dessicator dessicatorC.jpgwireN2c.jpg
Message ID: 12423     Entry time: Thu Aug 18 15:16:09 2016     In reply to: 12379     Reply to this: 12650
Author: Steve 
Type: Update 
Category: SUS 
Subject: SOS sus wire is in 

 

Stress Relieved 0.0017"  Music Wire  CFW P/N: CFW2035025,   Made 08-17-2016                                

    Old  2003    New  2016  
GBL 358.9 240.610 grams
UTS 357,061 229,603 PSI
YTS 343,211 177,371 PSI
ELONG 2.38 0.8 %
HEAT 10622 10622  

GBL (grams breaking load )         

UTS (ultimate tensile strength)                                 

YTS (yield tensile strength) 

ELONG (elongation)  

Quote:

0.0017" OD., 500ft steel music wire ordered. Pictures of the existing roll are below. It will be on 8" OD. spool too.

 

 

Attachment 1: 0.0017new.jpg  822 kB  | Show | Show all
ELOG V3.1.3-