Entry  Mon Jul 4 10:30:37 2016, ericq, Update, General, Preliminary vent plans 
    Reply  Tue Jul 5 17:14:02 2016, ericq, Update, General, Vent progress noOsemSide.JPGOSEMside.JPG
       Reply  Tue Jul 5 19:10:41 2016, rana, Update, General, Vent progress jumpEvent20151105.png
          Reply  Tue Jul 5 21:51:35 2016, ericq, Update, General, Vent progress 
             Reply  Tue Jul 5 23:38:42 2016, gautam, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation image.jpeg
                Reply  Wed Jul 6 22:58:01 2016, gautam, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation 
                   Reply  Thu Jul 7 00:25:07 2016, ericq, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation 
                      Reply  Thu Jul 7 10:49:03 2016, gautam, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation UL_glitchMon_Striptool.png
                         Reply  Fri Jul 8 11:35:45 2016, gautam, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation coilGlitchMon.png
                            Reply  Fri Jul 8 11:48:09 2016, Koji, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation 
                               Reply  Fri Jul 8 14:03:26 2016, ericq, Update, General, ETMX removed from suspension zoom.jpg
                                  Reply  Fri Jul 8 16:33:26 2016, ericq, Update, General, ETMX removed from suspension 
                                     Reply  Fri Jul 8 20:44:09 2016, Koji, Update, General, ETMX removed from suspension 6x
                                        Reply  Sun Jul 10 15:38:01 2016, ericq, Update, General, ETMX in guiderod/standoff gluing fixture 6x
                                           Reply  Tue Jul 12 16:08:00 2016, ericq, Update, General, ETMX ruby guided gluing 
                                              Reply  Wed Jul 13 21:13:51 2016, Koji, Update, General, ETMX guide rod gluing / ETMY Magnet gluing 13x
                                                 Reply  Thu Jul 14 22:29:57 2016, ericq , Update, General, ETMX guide rod gluing / ETMY Magnet gluing 
                                                    Reply  Fri Jul 15 16:23:51 2016, ericq , Update, General, Gluing setbacks IMG_2768.JPGIMG_2771.JPG
                                                       Reply  Tue Jul 19 16:39:29 2016, Lydia, Update, General, ETMX magnet gluing/guiderod excess glue removed P7190201.JPGP7190203.JPG
                                                          Reply  Wed Jul 20 13:58:55 2016, Steve, Update, SUS, SUS damping out of vac chamber ETMXrack.jpgfromETMX-satBox.jpgsusDampingSatboxCab.jpgsusDampingSatboxCabl.jpg
                                                          Reply  Thu Jul 21 17:01:59 2016, ericq, Update, General, ETMX magnet gluing/guiderod excess glue removed 
                                                             Reply  Fri Jul 22 23:03:23 2016, gautam, Update, General, ETMX suspension - progress IMG_2828.JPGIMG_2826.JPGIMG_2821.JPGIMG_2822.JPGIMG_2823.JPG
                                                                Reply  Sun Jul 24 03:43:56 2016, ericq, Update, General, Magnet positioning 
                                                                Reply  Mon Jul 25 13:01:49 2016, ericq, Update, General, ETMX suspension - progress IMG_2829.JPGIMG_2830.JPGIMG_2831.JPGIMG_2832.JPGIMG_2833.JPG
                                                                   Reply  Mon Jul 25 13:47:19 2016, rana, Update, General, ETMX suspension - progress 
                                                                   Reply  Mon Jul 25 17:01:54 2016, Steve, Update, SUS, ETMX suspension - progress ITMYside2011.JPG
                                                                   Reply  Mon Jul 25 20:37:46 2016, gautam, Update, General, ETMX suspension - Side OSEM test osem_test_2016_07_25.png
                                                                      Reply  Tue Jul 26 18:07:57 2016, gautam, Update, General, ETMX suspension - Further OSEM Tests OSEM_spectra.pdf
                                                                         Reply  Thu Jul 28 03:22:28 2016, ericq, Update, General, ETMX magnets re-reglued 
                                                                            Reply  Thu Jul 28 13:49:44 2016, gautam, Update, General, ETMX magnets re-reglued ETMX_sideMagnet.JPGETMX_sideMagnet_zoomed.JPG
                                                                               Reply  Thu Jul 28 17:48:17 2016, gautam, Update, General, ETMY LR magnet broke off 
                                                                                  Reply  Thu Jul 28 22:43:03 2016, rana, Update, General, ETMY UL magnet broke off 
                                                                                  Reply  Wed Aug 3 10:52:03 2016, Steve, Update, General, ETMY soaking dish SD1.jpgSD2.jpgSD3.jpg
                                                                               Reply  Fri Jul 29 03:32:30 2016, ericq, Update, General, ETMX suspended once more 
                                                                                  Reply  Fri Jul 29 13:17:34 2016, Koji, Update, General, Oven 
                                              Reply  Thu Jul 14 01:00:33 2016, Koji, Update, General, ETMX ruby guided gluing ruby.jpg
                                  Reply  Thu Jul 28 16:43:01 2016, Lydia, Update, General, ETMX aluminium standoff groove condition wire_damage_4.jpgwire_damage_zoom.jpgno_wire.jpgcomparison_2.jpg
                            Reply  Fri Jul 8 21:22:38 2016, gautam, Update, General, Vent progress - ETMX SUS Coil driver electronics investigation 
             Reply  Wed Jul 6 09:34:26 2016, Steve, Update, General, PRM sensing voltage LR PRM_LR.png
             Reply  Wed Jul 6 19:36:45 2016, Koji, Update, General, SUS Vmon HPF.png160706_ITMX_VMON2.pdf160706_ITMX_VMON1.pdf
                Reply  Thu Jul 7 15:23:39 2016, ericq, Update, General, SUS Vmon coil_drive.png
                   Reply  Thu Jul 7 16:05:55 2016, Koji, Update, General, SUS Vmon 
       Reply  Thu Jul 7 17:12:50 2016, Johannes, Update, General, Vent progress IMG_2621.JPGIMG_2623.JPGIMG_2628.JPGIMG_2595.JPG
          Reply  Tue Jul 12 23:51:16 2016, Johannes, Update, General, Vent progress - ETMY inspection IMG_2679.JPGIMG_2673.JPG
             Reply  Wed Jul 13 00:38:25 2016, Johannes, Update, General, Vent progress - ETMY attempted repositioning 
                Reply  Tue Jul 19 13:21:42 2016, Johannes, Update, General, Vent progress - ETMY attempted repositioning 
       Reply  Sun Jul 10 17:33:00 2016, ericq, Update, General, Vent progress 
          Reply  Mon Jul 11 15:13:22 2016, gautam, Update, General, Vent progress: in-date First Contact procured image.jpeg
Message ID: 12297     Entry time: Wed Jul 13 00:38:25 2016     In reply to: 12295     Reply to this: 12310
Author: Johannes 
Type: Update 
Category: General 
Subject: Vent progress - ETMY attempted repositioning 

[Lydia, Johannes]

We attempted to move the ETMY suspension near the access port in preparation for the cleaning process. The plan was to move in the face restraints first to the point of almost making contact, then the ones underneath so the optic is sitting on them, followed by the top one facing down, and then bringing in the stops on the faces.

While moving in the stoppers I noticed that the far lower stopper on the HR side was barely touching the face of the optic in its resting position and was basically pushing it sideways when moved forward. It was just on the edge, so I tried to compensate minimally by moving the underneath stops a little further on the near side, trying to let it 'slide' over a little so the screw would have better contact. I must have been too generous with the adjustment, because while proceeding I noticed at some point that the stick magnets on one side of the optic were not attached anymore but laying inside the OSEMs. The side magnet was also missing, it is now sitting on the suspension jig base plate. The dumbbells all seem intact, but we'll test them before we reglue the magnets to the optic. This is extremely unfortunate, but hopefully won't take too long to fix. At the very least, as Koji put it, the cleaning will be easier with the optic out of the suspension. Still, what a bummer.