Entry  Mon May 2 17:11:55 2016, rana, Update, COC, RC folding mirrors 
    Reply  Wed May 18 01:10:22 2016, gautam, Update, COC, Finesse modelling arms.pdfPRC.pdfSRC.pdfFinesse_model.zip
       Reply  Tue May 24 22:49:02 2016, gautam, Update, COC, Finesse modelling - mode overlap scans 9x
          Reply  Tue May 24 23:17:37 2016, ericq, Update, COC, Finesse modelling - mode overlap scans 
             Reply  Thu Jun 16 15:57:46 2016, gautam, Update, COC, Contrast as a function of RoC of ETMX contrastDefect.pdffinesseCode.zip
                Reply  Thu Jun 16 18:42:12 2016, rana, Update, COC, Contrast as a function of RoC of ETMX 
                   Reply  Thu Jun 16 23:02:57 2016, gautam, Update, COC, Contrast as a function of RoC of ETMX contrastDefect.pdf
                      Reply  Mon Jun 20 01:38:04 2016, rana, Update, COC, Contrast as a function of RoC of ETMX 
                         Reply  Mon Jun 20 18:07:15 2016, gautam, Update, COC, Contrast as a function of RoC of ETMX contrastDefectComparison.pdf
             Reply  Tue Jun 28 16:06:09 2016, gautam, Update, COC, RC folding mirrors - further checks C1_HOMcurves_Y.pdfC1_HOMcurves_DR.pdf
                Reply  Thu Jun 30 16:21:32 2016, gautam, Update, COC, Sideband HOMs resonating in arms image.jpegC1_HOMcurves_Y.pdfC1_HOMcurves_X.pdf
                Reply  Sat Aug 13 18:25:22 2016, gautam, Update, COC, RC folding mirrors - Numerical review PRX_consolidated.pdfSRX_consolidated.pdfGouy_PRC.pdfGouy_SRC.pdf
                   Reply  Tue Aug 16 11:51:43 2016, gautam, Update, COC, RC folding mirrors - Numerical review PRC_consolidated.pdfSRC_consolidated.pdfGouyPRC.pdfGouySRC.pdf
                      Reply  Tue Aug 16 16:38:00 2016, gautam, Update, COC, RC folding mirrors - Numerical review PRC_consolidated.pdfSRC_consolidated.pdfGouyPRC.pdfGouySRC.pdf
                         Reply  Wed Aug 17 14:37:36 2016, gautam, Update, COC, RC folding mirrors - Numerical review PRG.pdf
                            Reply  Wed Aug 17 16:28:46 2016, Koji, Update, COC, RC folding mirrors - Numerical review 
                            Reply  Mon Nov 21 15:34:24 2016, gautam, Update, COC, RC folding mirrors - updated specs Recycling_Mirrors_Specs_Nov2016.pdf
                               Reply  Thu Feb 23 10:59:53 2017, gautam, Update, COC, RC folding mirrors - coating optimization PR3_R_170222_2006.pdfPR3_123_TOnoise_170222_2203.pdfPR3_123_Layers_170222_2203.pdfPR3AR_R_170222_2258.pdfPR3AR_123_Layers_170222_2258.pdf
                                  Reply  Tue Mar 14 10:56:33 2017, gautam, Update, COC, RC folding mirrors - coating optimization PR3_R_170313_1701.pdfPR3AR_123_Layers_170313_1701.pdfPR3AR_R_170313_1752.pdfPR3AR_123_Layers_170313_1752.pdf
                                     Reply  Mon Apr 10 15:37:11 2017, gautam, Update, COC, RC folding mirrors - v3 of specs uploaded  8x
Message ID: 12204     Entry time: Mon Jun 20 18:07:15 2016     In reply to: 12197
Author: gautam 
Type: Update 
Category: COC 
Subject: Contrast as a function of RoC of ETMX 
Quote:

So, it seems that changing the ETMX for one of the spares will change the contrast defect from ~0.1% to 0.9%. True? Seems like that might be a big deal.

That is what the simulation suggests... I repeated the simulation for a PRFPMI configuration (i.e. no SRM, everything else  as per the most up to date 40m numbers), and the conclusion is roughly the same - the contrast defect degrades from ~0.1% to ~1.4%... So I would say this is significant. I also attempted to see what the contribution of the asymmetry in loss in the arms is, by running over the simulation with the current loss numbers of 230ppm for Yarm and 484ppm for the X arm, split equally between the ITMs and ETMs for both cases, and then again with lossless arms - see attachment #1. While this is a factor, this plot seems to suggest that the RoC mismatch effect dominates the contrast defect...

Attachment 1: contrastDefectComparison.pdf  73 kB  | Show | Show all