Mon Oct 19 11:20:34 2015, gautam, Update, VAC, RGA scan reset
|
Thu Apr 21 15:12:38 2016, Steve, Update, VAC, RGA is not working
|
Wed May 11 16:39:01 2016, ericq, Update, VAC, c0rga alive, output wonky
|
Thu May 12 14:29:58 2016, gautam, Update, VAC, RGA back up and running
|
Tue May 17 14:46:51 2016, Steve, Update, VAC, RGA scan at day 595 
|
Mon Aug 21 14:53:32 2017, gautam, Update, VAC, RGA reset
|
|
Message ID: 12086
Entry time: Thu Apr 21 15:12:38 2016
In reply to: 11697
Reply to this: 12115
|
Author: |
Steve |
Type: |
Update |
Category: |
VAC |
Subject: |
RGA is not working |
|
|
Quote: |
Steve pointed out that in the aftermath of the Nitrogen running out a couple of times last week, the RGA had shut itself off thinking that there was a leak and so it was not performing the scheduled scans once a day. So the data files from the scheduled scans were empty in the /opt/rtcds/caltech/c1/scripts/RGA/logs directory. The wiki page for getting it up and running again is up-to-date, but the script RGAset.py did not exist on the c0rga machine, which the RGA is communicating with via serial port. I copied over the script RGAset.py from rossa to c0rga and ran the script on that machine - but the error flags it returned were not all 0 (indicating some error according to the manual) - so I edited the script to send just the initialize command ('IN0') and commented out the other commands, after which I got error flags which were all 0. After this, I ran a manual scan using 'RGAlogger.py', and it appears that the RGA is now able to take scans again - I'm attaching a plot of the scan results. We've saved this scan as a reference to compare against after a few days.
|
Our last RGA scan is from February 14, 2016 We had a power outage on the 15th
Gautom has not succeded reseting it. The old c0rga computer looks dead. Q may resurrect it, if he can? |