40m
QIL
Cryo_Lab
CTN
SUS_Lab
TCS_Lab
OMC_Lab
CRIME_Lab
FEA
ENG_Labs
OptContFac
Mariner
WBEEShop
|
40m Log |
Not logged in |
 |
|
Tue Jan 26 17:50:22 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Wed Jan 27 18:19:45 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Thu Jan 28 00:29:30 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement 
|
Mon Feb 1 18:11:25 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Fri Jan 29 14:43:47 2016, Steve, Update, Green Locking, Innolight laser is 10 years old
|
Fri Jan 29 16:55:27 2016, rana, Update, Green Locking, Innolight laser is 10 years old
|
Sat Jan 30 00:12:22 2016, gautam, Update, Green Locking, Innolight laser is 10 years old
|
Sat Jan 30 09:56:24 2016, Koji, Update, Green Locking, Innolight laser is 10 years old
|
Mon Feb 1 15:16:28 2016, gautam, Update, Green Locking, Innolight laser is 10 years old
|
Mon Feb 1 15:43:18 2016, Koji, Update, Green Locking, Innolight laser is 10 years old
|
Fri Feb 5 15:02:13 2016, gautam, Update, Green Locking, X-end NE cable
|
Fri Feb 12 11:10:49 2016, Steve, Update, Green Locking, Innolight laser is 10 years old
|
Mon Mar 21 14:29:32 2016, Steve, Update, Calibration-Repair, 1W Innolight laser repair diagnoses
|
Thu Mar 24 07:56:09 2016, Steve, Update, Calibration-Repair, NO Noise Eater for 1W Innolight
|
Mon Apr 11 17:03:41 2016, Steve, Update, Calibration-Repair, 1W Innolight repair completed
|
|
Message ID: 12040
Entry time: Mon Mar 21 14:29:32 2016
In reply to: 11961
Reply to this: 12045
|
Author: |
Steve |
Type: |
Update |
Category: |
Calibration-Repair |
Subject: |
1W Innolight laser repair diagnoses |
|
|
Quote: |
Quote: |
After adjusting the alignment of the two beams onto the PD, I managed to recover a stronger beatnote of ~ -10dBm. I managed to take some measurements with the PLL locked, and will put up a more detailed post later in the evening. I turned the IMC autolocker off, turned the 11MHz Marconi output off, and closed the PSL shutter for the duration of my work, but have reverted these to their nominal state now. The are a few extra cables running from the PSL table to the area near the IOO rack where I was doing the measurements from, I've left these as is for now in case I need to take some more data later in the evening...I
|
Innolight 1W 1064nm, sn 1634 was purchased in 9-18-2006 at CIT. It came to the 40m around 2010
It's diodes should be replaced, based on it's age and performance.
RIN and noise eater bad. I will get a quote on this job.
The Innolight Manual frequency noise plot is the same as Lightwave' elog 11956
|
Diagnoses from Glasglow:
“So far we have analyzed the laser. The pump diode is degraded. Next we would replace it with a new diode. We would realign the diode output beam into the laser crystal. We check all the relevant laser parameters over the whole tuning range. Parameters include single direction operation of the ring resonator, single frequency operation, beam profile and others. If one of them is out of spec, then we would take actions accordingly. We would also monitor the output power stability over one night. Then we repackage and ship the laser.” |