Tue Jan 26 17:50:22 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Wed Jan 27 18:19:45 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Thu Jan 28 00:29:30 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement 
|
Mon Feb 1 18:11:25 2016, gautam, Update, Green Locking, Lightwave frequency noise measurement
|
Fri Jan 29 14:43:47 2016, Steve, Update, Green Locking, Innolight laser is 10 years old
|
Fri Jan 29 16:55:27 2016, rana, Update, Green Locking, Innolight laser is 10 years old
|
Sat Jan 30 00:12:22 2016, gautam, Update, Green Locking, Innolight laser is 10 years old
|
Sat Jan 30 09:56:24 2016, Koji, Update, Green Locking, Innolight laser is 10 years old
|
Mon Feb 1 15:16:28 2016, gautam, Update, Green Locking, Innolight laser is 10 years old
|
Mon Feb 1 15:43:18 2016, Koji, Update, Green Locking, Innolight laser is 10 years old
|
Fri Feb 5 15:02:13 2016, gautam, Update, Green Locking, X-end NE cable
|
Fri Feb 12 11:10:49 2016, Steve, Update, Green Locking, Innolight laser is 10 years old
|
Mon Mar 21 14:29:32 2016, Steve, Update, Calibration-Repair, 1W Innolight laser repair diagnoses
|
Thu Mar 24 07:56:09 2016, Steve, Update, Calibration-Repair, NO Noise Eater for 1W Innolight
|
Mon Apr 11 17:03:41 2016, Steve, Update, Calibration-Repair, 1W Innolight repair completed
|
|
Message ID: 11963
Entry time: Sat Jan 30 00:12:22 2016
In reply to: 11962
Reply to this: 11964
|
Author: |
gautam |
Type: |
Update |
Category: |
Green Locking |
Subject: |
Innolight laser is 10 years old |
|
|
Quote: |
I don't think there's any evidence that the noise eater is bad. That would change the behavior of the relaxation oscillation which is at 1 MHz ?
|
While I was investigating the AM/PM ratio of the Innolight, I found that there was a pronounced peak in the RIN at ~400kHz, which did not change despite toggling the noise eater switch on the front panel (see plot attached). The plot in the manual suggests the relaxation oscillations should be around 600kHz, but given that the laser power has dropped by a factor of ~3, I think it's reasonable that the relaxation oscillations are now at ~400kHz? |
|
|