40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Dec 17 01:55:16 2015, ericq, Update, LSC, Uncooperative AUX X 
    Reply  Fri Dec 18 17:37:04 2015, rana, Update, LSC, Uncooperative AUX X 
       Reply  Mon Dec 21 02:29:49 2015, ericq, Update, LSC, AUX X RIN measurements auxXRIN.pdf
          Reply  Tue Jan 5 02:54:38 2016, ericq, Update, LSC, AUX X Freq Noise attempt 
             Reply  Tue Jan 5 13:17:06 2016, rana, Update, LSC, AUX X Freq Noise attempt 
                Reply  Tue Jan 5 16:33:45 2016, ericq, Update, LSC, AUX X Freq Noise attempt 
                   Reply  Thu Jan 7 04:28:39 2016, ericq, Update, LSC, AUX X Freq Noise measured PLLspec.pdfPLL_OLG.pdf
                      Reply  Thu Jan 7 16:52:32 2016, ericq, Update, LSC, AUX X Freq Noise measured AUXfreqnoise.pdfauxXmeasurements.zip
                         Reply  Thu Jan 7 19:04:25 2016, Koji, Update, LSC, AUX X Freq Noise measured 
                            Reply  Fri Jan 8 14:47:33 2016, ericq, Update, LSC, AUX Y Freq Noise measured AUX_freqnoises.pdfAUXY_Jan8.zip
                               Reply  Fri Jan 8 20:02:49 2016, rana, Update, LSC, AUX Y Freq Noise measured 
                               Reply  Sat Jan 9 00:39:15 2016, gautam, Update, LSC, AUX Y Freq Noise re-measured AuxPLL.pdf
                                  Reply  Mon Jan 11 19:01:56 2016, gautam, Update, LSC, PLL Marconi Investigation 6x
                                     Reply  Tue Jan 12 03:03:55 2016, ericq, Update, LSC, Frequently making noise 
                                        Reply  Tue Jan 12 19:38:31 2016, gautam, Update, LSC, Frequently making noise 2016_01_AUXLaser.tar.gzOLGs.pdfvariedCurrent.pdfvariedTemp.pdfPLL_setup.pdf
Message ID: 11892     Entry time: Fri Dec 18 17:37:04 2015     In reply to: 11889     Reply to this: 11894
Author: rana 
Type: Update 
Category: LSC 
Subject: Uncooperative AUX X 

Here's how we should diagnose the EX laser:

  1. Compare IR RIN of laser out to 100 kHz with that of another similar NPRO.
  2. Look at time series of IR beat signal with a fast scope. Are there any high frequency glitches?
  3. Disconnect all of the cables to the EX laser PZT and temperature control. Does the frequency noise change?
  4. Change the temperature by +/- 1 deg to move away from mode hop regions. Remeasure RIN and frequency noise and plot.