Entry  Wed Dec 2 17:14:33 2015, yutaro, Update, LSC, Beam on POX11 is possibly not centered well 47.png41.png
    Reply  Fri Dec 4 12:33:52 2015, yutaro, Update, LSC, Beam on POX11 is possibly not centered well image1-3.JPG07.png47.png
       Reply  Fri Dec 4 23:02:13 2015, yutaro, Update, LSC, Beam on POX11 is possibly not centered well image1.JPG56.png04.png
Message ID: 11839     Entry time: Wed Dec 2 17:14:33 2015     Reply to this: 11847
Author: yutaro 
Type: Update 
Category: LSC 
Subject: Beam on POX11 is possibly not centered well 

I checked how POXDC level changes when the angle of ITMX is varied. ETMX was misaligned.

Then I found that in YAW direction the POXDC level is maximized but it doesnt have plateau, and in PIT direction it is not maximized so that it is at the slope and it doesnt have plateau, as shown in attached figures. These results indicate that the beam size on POX11 is not small enough compared to the size of the diode and it is not centered well.

Attachment 1: 47.png  12 kB  | Hide | Hide all
47.png
Attachment 2: 41.png  12 kB  | Hide | Hide all
41.png