Message ID: 11821     Entry time: Sun Nov 29 05:23:57 2015     In reply to: 11817     Reply to this: 11822
Author: rana 
Type: Update 
Category: LSC 
Subject: Current state of the frequency source, and possible improvement 

I need some more hints to understand the improvement, although its generally good to re-build it considering the sad state of the assembly/installation that you found.

I see that the current design brings the 11 MHz signal to -2 dBm before intering the first ZHL-2+, but since that has a NF of 9 dB, that seems to only degrade the phase noise to -2 - (-174 +9) = -163 dBc. That seems OK since we only need -160 dBc from this system. Probably the AM noise is worse than this already (we should remember to hook up a simple AM stabilizer in 2016, as well as the ISS).

What else are the main features of this improvement? I can reward a good summary with some Wagonga.