Entry  Wed Nov 25 16:40:32 2015, yutaro, Update, LSC, round trip loss of Y arm 
    Reply  Wed Nov 25 23:34:52 2015, yutaro, Update, LSC, round trip loss of Y arm 14.png
       Reply  Fri Nov 27 03:38:23 2015, yutaro, Summary, LSC, round trip loss of Y arm image1.png
          Reply  Fri Nov 27 22:20:24 2015, yutaro, Update, LSC, round trip loss of Y arm log.txt.zip17.png
          Reply  Sat Nov 28 11:46:40 2015, yutaro, Update, LSC, possible error source of loss map measurement 
          Reply  Mon Nov 30 10:41:45 2015, yutaro, Update, LSC, Does a baffle in front of ETMY have effect on loss map measurement? 
             Reply  Mon Nov 30 17:17:30 2015, yutaro, Update, LSC, Does a baffle in front of ETMY have effect on loss map measurement? element2-2.pngelement0-0.pngelement4-4.png
          Reply  Mon Dec 7 11:11:25 2015, yutaro, Summary, LSC, round trip loss of X arm image1-2.JPGsymmetry.png
Message ID: 11818     Entry time: Fri Nov 27 03:38:23 2015     In reply to: 11816     Reply to this: 11819   11820   11823   11857
Author: yutaro 
Type: Summary 
Category: LSC 
Subject: round trip loss of Y arm 

Tonight I measured "loss map" of ETMY. The method to calculate round trip loss is same as written in elog 11810, except that I used POYDC instead of ASDC this time.

How I changed beam spot on ETMY is: elog 11779.

I measured round trip loss for 5 x 5 points. The result is below.

(unit: ppm)

494.9 +/- 7.6       356.8 +/- 6.0       253.9 +/- 7.9       250.3 +/- 8.2       290.6 +/- 5.1      
215.7 +/- 4.8       225.6 +/- 5.7       235.1 +/- 7.0       284.4 +/- 5.4       294.7 +/- 4.5      
205.2 +/- 6.0       227.9 +/- 5.8       229.4 +/- 7.2       280.5 +/- 6.3       320.9 +/- 4.3      
227.9 +/- 5.7       230.5 +/- 5.5       262.1 +/- 5.9       315.3 +/- 4.7       346.8 +/- 4.2      
239.7 +/- 4.5       260.7 +/- 5.3       281.2 +/- 5.8       333.7 +/- 5.0       373.8 +/- 4.9 

The correspondence between the loss shown above and the beam spot on ETMY is shown in the following figure. In the figure, "downward" and "left" indicate direction of shift of the beam spot when you watch it via the camera (ex. 494.9 ppm corresponds to the lowest and rightest point). 

Edited below on 28th Nov. 

To shift the beam spot on ETMY, I added offset in YARM dither loop. The offset was [-30,-15,0,15,30]x[-10,-5,0,5,10] for pitch and yaw, respectively.  How I calibrated the beam spot is basically based on elog 11779, but I multiplied 5.3922 for vertical direction and 4.6205 for horizontal direction which I had obtained by caliblation of oplev (elog 11785).

Edited above on 28 th Nov.    

 

   

 

I will report the detail later. 