Message ID: 11781
Entry time: Wed Nov 18 16:53:17 2015
In reply to: 11780
Reply to this: 11782
|
Author: |
Koji |
Type: |
Summary |
Category: |
PSL |
Subject: |
PMC PZT capacitance |
|
|
The PMC PZT capacitance was measured.
- Turn off the HV supplies. Disconnect HV OUT cable.
- Make sure the cable is discharged.
- Measure the capacity at the cable end with an impedance meter.
=> The PMC PZT capacitance at the cable end was measured to be 222nF
Combined with the output impedance of 63.3kOhm, the LPF pole is at 11.3Hz |