Message ID: 11758     Entry time: Thu Nov 12 10:47:17 2015     In reply to: 11752
Author: ericq 
Type: Update 
Category: SUS 
Subject: ETMX oplev servo disabled 

By looking at the oplev and Vmons before and after a step of 90 counts in SUS-ETMX_PIT_OFFSET, I observe:

  • UL: +1.53mV / urad
  • UR: -1.94mV / urad
  • LR: -1.54mV / urad
  • LL: +1.92mv / urad

The random error associated with these measurements is ~0.02mV

So, the ~7urad urad shift seen in my earlier post would mean a change of around 10mV in the Vmon signals, which isn't evident in the traces. So, this is possibly a piece of evidence in favor of a real mechanical shift, rather than an electronic glitch. 