I found (an old) 10 dB coupler in the RF component shelves near MC2 - it has BNC connectors and not SMA connectors, but I thought it would be worth it to switch out the 20dB coupler currently on the X green beat PD on the PSL table with it. I used some BNC to SMA adaptors for this purpose. It appears that the coupler works, because I am now able to register an input signal on the X arm channel of the digital frequency counter (i.e. the coupled output from the green beat PD). I thought it may be useful to have this in place and do an IR transmissions arm scan using ALS for the X arm as well, in order to compare the results with those discussed here. However, the beat note amplitude on the analyzer in the control room looks noticeably lower - I am not sure if the coupler is responsible for this or if it has to do with the problems we have been having with the X end laser (the green transmission doesn't look glitchy on striptool though, and the transmission itself is ~0.45). In any case, we could always remove the coupler if this is hindering locking efforts tonight. |