The new ZKL-1R5 RF amplifier that Steve ordered arrived yesterday. I installed this in the frequency divider box and did a quick check using the Fluke RF signal generator and an oscilloscope to verify that both the X and Y paths were working.
I've now installed the box in the 1X2 rack where the olf "RF amplifiers for ALS and FOL" box used to sit (I swapped that out as I needed the L brackets on that chassis to mount mine, see Attachment #1 for the new layout). The power cable that used to power the old chassis was available, but the connector was of the wrong gender, so I had to switch this out. After verifying that I was getting the correct voltage (+15V), I connected it to the chassis.
I then did a quick check with the Fluke generator to make sure that all was working as expected - Eric had set up some ADC channels for me earlier today in the C1ALS model, and I copied over my frequency counting module from C1TST into C1ALS, and recompiled the model. The RF generator was set to generate a 25MHz signal at -20dBm, which I then split using an RF power splitter between the X and Y arms. I then checked the output using dataviewer - I recovered an output frequency of ~27.64 MHz with a jitter of ~0.02 MHz with a 20Hz low-pass filter in place (see Attachment #2), which looks consistent with the systematic error inherent in the zero-crossing counting algorithm and random fluctuations I had observed in my earlier trials, discussed here. But a more systematic investigation needs to be carried out in this regard. The interfacing between the hardware and software seems to be working alright though. I've left the RF generator near the 1x2 rack for now, though its powered off.
The mode cleaner unlocked quite a few times while I was working but looks stable now.
|