Message ID: 11680     Entry time: Fri Oct 9 14:50:18 2015     In reply to: 11679     Reply to this: 11681
Author: Koji 
Type: Update 
Category: LSC 
Subject: ALS plant shape 

ALS is the comparison of the PSL laser freq vs the end laser freq that is locked to the arm cavity resonant freq

On the other hand, the AS55 PDH is the comparison of the PSL laser freq after the IMC vs the arm cavity resonant freq. Here the PDH signal involves the arm cavity pole.

In total you observe the difference by the IMC cav pole + the arm cav pole.