1. POP110 RF amps are powered from the cross connect. But that +15V block has GND connections that are not connected to the ground.
i.e. The ground potential is given by the signal ground. (Attachment 1)
This is caused by the misuse of the DIN connector blocks. The hod side uses an isolated block assuming a fuse is inserted.
However, the ground sides also have the isolated blocks
2. One of the POP110 RF cable has a suspicious shiled. The rigidity of the cable is low, suggesting the broken shield. (Attachment 2) |