40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Jul 16 03:01:57 2015, ericq, Update, LSC, Old beatbox hooked back up beat_comparison.pngaLIGO_vs_beatbox.xml.zip
    Reply  Thu Jul 23 03:09:05 2015, ericq, Update, LSC, aLIGO demod board lives! beatbox_vs_aLIGO.png
       Reply  Thu Jul 30 14:42:03 2015, ericq, Update, LSC, aLIGO demod board lives! partialALSbudget.pngdemodDriveLevels.pngALScoherences.pngpartialALSbudget.zip
          Reply  Thu Jul 30 15:58:00 2015, ericq, Update, LSC, Beat note Alignment fluctuation effects measured RF_RIN.pngRF_RINspec.pngRFampCoh.xml.zip
             Reply  Tue Aug 4 03:02:30 2015, ericq, Update, LSC, Beat note Alignment fluctuation effects measured coherences.pnglinX.png
                Reply  Tue Aug 11 02:40:29 2015, ericq, Update, LSC, 50m delay lines - Rough calibrations 
Message ID: 11490     Entry time: Tue Aug 11 02:40:29 2015     In reply to: 11478
Author: ericq 
Type: Update 
Category: LSC 
Subject: 50m delay lines - Rough calibrations 

Jessica will soon ELOG about some measurements suggesting that the conductive connector-ized ALS delay line enclosure is the way to go, when considering crosstalk between the delay lines. It is currently mounted and hooked up on the LSC rack, though I need to make a bunch of new SMA cables now that I think a semi-permanent arrangement has been reached. 

I did a rough re-calibration of the phase tracker output, since the increased cable delay changes the degree/Hz gain. This was done by fitting a line to a slow sawtooth FM of the SRS DS345's (1Hz rate, 10kHz deviation, 30MHz carrier). This resulted in the following calibration updates

  • ALSX: 19230 -> 13051 Hz/count, 3.4dB more sensitive

  • ALSY: 19425 -> 12585 Hz/count, 3.8db more sensitive

Again, this is a rough calibration. Nevertheless, it is not so surprising we don't get the 50m/30m = 4.4dB increase we would expect just from the lengths; the (I presume) increased cable loss matters. Also, the loss' frequency dependance is an additional reason that the phase tracker calibration is not constant over all frequencies. 

I took spectra with the arms in IR lock, but didn't see any real improvement beyond a possible dip in the floor from 100-200Hz. This doesn't surprise me too much, however, since I don't believe that we are currently dominated by electronic noises that this gain increase would help overcome. 

Last week, Koji mentioned the ALS phase noise added due to the post-cavity table motion the arm-transmitted green beams experience before hitting the beat PD. I should estimate the size of this effect for our situation. 