Message ID: 11477     Entry time: Mon Aug 3 18:19:09 2015
Author: Jessica 
Type: Update 
Category: General 
Subject: Anodization of front panels accounted for 

Previously, I had gotten the same results for the conductive and the isolated front panels. Today, I sanded off the anodized part on the back of the conductive front panel. I checked afterwards with a mulitmeter to ensure that it was indeed conductive through all the SMA connectors. 

I drove a frequency of 29.359 Hz through the X Arm cable and 29.3592 Hz through the Y Arm cable, giving a difference of 200 Hz. Previously, there would only be a spike in the Y Arm at the difference, while the X Arm did not change if the Y arm was on or off. Now that the panel is fully conductive, a spike can also be seen in the X arm, indicating that crosstalk may possibly be happening with this panel, now that the spike corresponds to both the X arm and Y arm. These results are only after one set of data. Tomorrow I'll take two more sets of data with this panel and do a more in depth comparison of these results to what had been previously seen.  

Attachment 1: redo_conduct1X.png  133 kB  | Hide | Hide all
redo_conduct1X.png
Attachment 2: redo_conduct1Y.png  145 kB  | Hide | Hide all
redo_conduct1Y.png