40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri May 22 18:09:58 2015, ericq, Update, Computer Scripts / Programs, ifoCoupling ETMX_PIT_L_coupling.png
    Reply  Tue May 26 19:57:11 2015, rana, Update, Computer Scripts / Programs, ifoCoupling 
Message ID: 11325     Entry time: Tue May 26 19:57:11 2015     In reply to: 11321
Author: rana 
Type: Update 
Category: Computer Scripts / Programs 
Subject: ifoCoupling 

Looks like a very handy code, especially with the real statistical tests.

I would make sure to use much smaller excitation amplitudes. Since the coupling is nonlinear, we expect that its only a good noise budget estimator when the excitation amplitude is less than a factor of 3 above the quiesscent excitation.

ELOG V3.1.3-