Fri Dec 26 17:45:21 2014, Steve, Update, VAC, vac pressure rose to 1.3 mTorr 
|
Sun Dec 28 07:29:16 2014, Steve, Update, VAC, vacuum is normal
|
Tue Feb 3 15:20:13 2015, Steve, Koji, Update, VAC, Pneumatic pressure is being read again
|
Sun Mar 22 18:42:40 2015, Steve, Update, VAC, vac pressure rose to 1.2mTorr 
|
Mon Mar 23 10:36:55 2015, ericq, Update, VAC, Pressure watch script
|
Fri Apr 24 17:30:32 2015, Jenne, Update, VAC, Pressure watch script broken
|
Sat Apr 25 18:50:47 2015, ericq, Update, VAC, Pressure watch script broken
|
Sat Apr 25 22:17:49 2015, rana, Update, CDS, MXstream restart script working (beta)
|
Tue May 19 17:03:34 2015, Koji, Update, CDS, MXstream restart script working (beta)
|
Tue May 19 18:38:33 2015, rana, Update, CDS, MXstream restart script working (beta)
|
Mon Apr 27 14:37:55 2015, Steve, Update, VAC, N2 pneumatic pressure watch
|
Thu Apr 30 16:30:25 2015, Steve, Update, VAC, N2 pneumatic pressure watch set up
|
Tue Jan 31 14:14:29 2017, rana, Update, VAC, Pressure watch script
|
Sun May 15 19:48:08 2016, Steve, Update, VAC, run out of N2
|
Tue Jun 14 10:13:34 2016, Steve, Update, VAC, temporary N2 supply line
|
Fri Jun 17 15:22:31 2016, gautam, Update, VAC, N2 supply line restored after retiling
|
|
Message ID: 11264
Entry time: Thu Apr 30 16:30:25 2015
In reply to: 11260
|
Author: |
Steve |
Type: |
Update |
Category: |
VAC |
Subject: |
N2 pneumatic pressure watch set up |
|
|
We have 2 transduser PX303-3KG5V http://www.omega.com/pressure/pdf/PX303.pdf They will be installed on the out put of the N2 cylinders to read the supply pressure.
I will order one DC power supply http://www.omega.com/pptst/PSU93_FPW15.html PSU-93
One full cylinder pressure is ~ 2400 PSI max so two of them will give us ~9Vdc
The email reminder should be send at 1000 PSI = 1.8 V
|