40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Dec 26 17:45:21 2014, Steve, Update, VAC, vac pressure rose to 1.3 mTorr noN2.pngbacktoVacNormal.png
    Reply  Sun Dec 28 07:29:16 2014, Steve, Update, VAC, vacuum is normal 3d.png
    Reply  Tue Feb 3 15:20:13 2015, Steve, Koji, Update, VAC, Pneumatic pressure is being read again N2pressureReadingIsBack.png
    Reply  Sun Mar 22 18:42:40 2015, Steve, Update, VAC, vac pressure rose to 1.2mTorr NoN2.pngPressureReadingWorks.png
       Reply  Mon Mar 23 10:36:55 2015, ericq, Update, VAC, Pressure watch script 
          Reply  Fri Apr 24 17:30:32 2015, Jenne, Update, VAC, Pressure watch script broken 
             Reply  Sat Apr 25 18:50:47 2015, ericq, Update, VAC, Pressure watch script broken 
                Reply  Sat Apr 25 22:17:49 2015, rana, Update, CDS, MXstream restart script working (beta) 
                   Reply  Tue May 19 17:03:34 2015, Koji, Update, CDS, MXstream restart script working (beta) 
                      Reply  Tue May 19 18:38:33 2015, rana, Update, CDS, MXstream restart script working (beta) 
          Reply  Mon Apr 27 14:37:55 2015, Steve, Update, VAC, N2 pneumatic pressure watch  
             Reply  Thu Apr 30 16:30:25 2015, Steve, Update, VAC, N2 pneumatic pressure watch set up 
          Reply  Tue Jan 31 14:14:29 2017, rana, Update, VAC, Pressure watch script 
       Reply  Sun May 15 19:48:08 2016, Steve, Update, VAC, run out of N2 noN2.png
          Reply  Tue Jun 14 10:13:34 2016, Steve, Update, VAC, temporary N2 supply line afterN2work.png
             Reply  Fri Jun 17 15:22:31 2016, gautam, Update, VAC, N2 supply line restored after retiling c0vac_06172016.png
Message ID: 11260     Entry time: Mon Apr 27 14:37:55 2015     In reply to: 11159     Reply to this: 11264
Author: Steve 
Type: Update 
Category: VAC 
Subject: N2 pneumatic pressure watch  

 

Quote:

Based on Jenne's chiara disk usage monitoring script, I made a script that checks the N2 pressure, which will send an email to myself, Jenne, Rana, Koji, and Steve, should the pressure fall below 60psi. I also updated the chiara disk checking script to work on the new Nodus setup. I tested the two, only emailing myself, and they appear to work as expected. 

The scripts are committed to the svn. Nodus' crontab now includes these two scripts, as well as the crontab backup script. (It occurs to me that the crontab backup script could be a little smarter, only backing it up if a change is made, but the archive is only a few MB, so it's probably not so important...)

This watch script gives little time to replace N2 cylinder. When the regulated supply drops below 60 psi the cylinder pressure is 60 psi too.

It is more of a statement that V1 is closed and act accordingly. It's only practical if you are in the lab.

Rana pointed it out correctly that we need this message 24 hrs before it happens. This requires monitoring  of total supplies , not the regulated one.

So we need pressure transducers on each nitrogen cylinder, before the regulator. The sum of the two N2 cylinder when they are full 4000 - 4500 psi

The first email should be send out at 1000 psi as sum of the two cylinders. This means that you have  1 day to replace nitrogen cylinder.

 Most of the time the daily consumption is 750 +-50 psi   

However sometimes  this variation goes up   ~750 +-150 psi