Message ID: 1125     Entry time: Mon Nov 10 11:06:09 2008
Author: rob 
Type: HowTo 
Category: IOO 
Subject: mode cleaner locked 

I found the mode cleaner unlocked, with (at least) MC1 badly mis-aligned. After checking the coil alignment biases and finding everything there looking copasetic, I checked the trends of SUS{PIT,YAW,POS} and found that both MC1 and MC3 took a step this morning. The problem turned out to be loosed/jiggled cables at the satellite amplifiers for these suspensions. Giving them a good hard push to seat them restored the alignment and the mode cleaner locked right up.