40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Mar 5 12:27:27 2015, manasa, Update, General, FOL fiber box on the PSL table 
    Reply  Fri Mar 6 14:29:59 2015, manasa, Update, General, FOL troubleshooting 
       Reply  Mon Mar 9 14:43:31 2015, manasa, Update, General, FOL troubleshooting FOL_trouble.pdfIMAG0341.jpgIMAG0342.jpg
          Reply  Tue Mar 10 17:48:07 2015, manasa, Update, General, FOL troubleshooting 
             Reply  Wed Mar 11 19:48:25 2015, Koji, Update, General, FOL troubleshooting 
                Reply  Thu Mar 12 11:57:38 2015, Koji, Update, General, FOL troubleshooting 
Message ID: 11123     Entry time: Mon Mar 9 14:43:31 2015     In reply to: 11110     Reply to this: 11128
Author: manasa 
Type: Update 
Category: General 
Subject: FOL troubleshooting 

Figuring out the problem with frequency counter readouts:

RF amplifier box:
The frequency counter was setup to read the beat frequency after amplification of the RFPD signals. The output as seen on the frequency counter screen as well as the epics values was intermittent. Also, locking the arms and changing the end laser frequencies did not change the frequency counter outputs. It looks like the RF amplifier is oscillating (Oscillating frequencies were ~107 MHz and 218 MHz) and the input circuit needs to be modified and the connections need better insulation.

As of now, I have connected the RFPD outputs to the frequency counter sans any amplification and we are able to get a readout of the beat frequency at > 40MHz (Frequency counter requires much higher amplitude signals for frequencies lower than that).

The frequency counter has 4 ranges of operation: 
Range 1 : 1 - 40MHz
Range 2: 40 - 190MHz
Range 3: 190 - 1400MHz  
Range 4: 1400 - 6000MHz

I set up the beat frequency readout in various configurations to troubleshoot and have recorded my observation (attachments). The amplifiers in all cases is just a single ZFL-500LN.

There seems to be a problem with the RF amplifier and the frequency counter when they are set together. I am not able to figure it out and stuck right here sad

Attachment 1: schematic for readout and corresponding observations

Attachment 2: Oscilloscope screen snapshot for schematic 3

Attachment 3: Spectrum analyzer screen snapshot for schematic 3

More info: RFPD used is Thorlabs FPD310 and frequency counter is Mini Circuits UFC-6000. The RF amplifier has decoupling capacitors soldered across the power supply.

Attachment 1: FOL_trouble.pdf  260 kB  | Hide | Hide all | Show all
FOL_trouble.pdf
Attachment 2: IMAG0341.jpg  1.783 MB  | Show | Hide all | Show all
Attachment 3: IMAG0342.jpg  2.364 MB  | Hide | Hide all | Show all
IMAG0342.jpg
ELOG V3.1.3-