Message ID: 11054     Entry time: Fri Feb 20 12:29:01 2015     In reply to: 11053
Author: ericq 
Type: Update 
Category: CDS 
Subject: All optics damped 

I noticed some diagnostic bits in the c1sus IOP complaining about user application timing and FIFO under/overflow (The second and fourth squares next to the DACs on the GDS screen.) Over in crackle-cymac land, I've seen this correlate with large excess DAC noise. After restarting all models, all but one of these is green again, and the optics are now all damped. 

It seems there were some fishy BURT restores, as I found the TT control filters had their inputs and outputs switched off. Some ASS filters were found this way too. More undesired settings may still lurk in the mists...

The interferometer is now realigned, arms locked. 