Message ID: 11008     Entry time: Thu Feb 12 01:00:18 2015     In reply to: 11004
Author: rana 
Type: Update 
Category: LSC 
Subject: RFPD spectra 

The nonlinearity in the LSC detection chain (cf T050268) comes from the photodetector and not the demod board. The demod board has low pass or band pass filters which Suresh installed a long time ago (we should check out what's in REFL33 demod board). 

Inside the photodetector the nonlinearity comes about because of photodiode bias modulation (aka the Grote effect) and slew rate limited distortion in the MAX4107 preamp.