Entry  Fri Dec 12 04:33:49 2014, Jenne, Update, ASC, ASS retuned 
    Reply  Wed Dec 17 01:51:44 2014, rana, jenne, Update, ASC, ASS retuned ASSconfig_141217_0205.png
       Reply  Wed Dec 17 19:04:12 2014, jenne, Update, ASC, ASS retuned 
          Reply  Wed Dec 17 19:31:55 2014, Koji, Update, ASC, ASS retuned 
             Reply  Tue Jan 27 21:33:39 2015, Koji, Update, ASC, ASS retuned 
Message ID: 10946     Entry time: Tue Jan 27 21:33:39 2015     In reply to: 10813
Author: Koji 
Type: Update 
Category: ASC 
Subject: ASS retuned 

I checked the situation of ASS. I wanted to know how much we are away from the maximum transmittion.

Conclusion:
ASS makes the X arm shifted from the maximum transmission. This causes the contrast degraded by ~3%.
We need to fix the Xarm ASS so that it can maximize the transmission and ignor the spot centering at ITMX.


Conditioning before the measurement:

- ASDC offset was removed
- X&Y arm was aligned by ASS

With ASS:

Average transmission: 0.86
Pmax = 1045 +/- 9 cnts
Pmin = 22 +/- 4 cnts

==> Contrast = (Pmax - Pmin)/(Pmax+Pmin) = 0.960+/-0.007

After manual alignment of the X arm (ignoring spot centering):

Average transmission: 0.88
Pmax = 1103 +/- 11 cnts
Pmin = 5 +/- 1 cnts

==> Contrast = (Pmax - Pmin)/(Pmax+Pmin) = 0.991+/-0.002